Features
○ Integrated breathing circuit with APL valve, excellent tightness.
○ Replacing drive bellow is NOT required when operating adult/pediatric anesthesia.
○ Applied with 2 high-precision vaporizers. (Selectatec).
○ 10.1 inch touch screen with no dead space.
○ Real-time detection for loops, Paw, BPM, VT and compliance.
○ Comprehensive monitoring capabilities include ventilation parameters monitoring,preset parameters, alarm indication and breathing waveform monitoring.
○ Head light.
○ Altitude compensation function.
○ ACGO available to connect with T/Bain circuit for open operation.
○ Yoke system with up to 3 back up cylinders (A type with 11 L size) (Optional).
Technique Specifications | |
Physical Specifications | |
Dimensions | 70*85*140cm |
Weight | 65kg |
Casters | Front wheels with lock |
Screen | 10.1 inch LCD tochscreen |
Pipeline Supply | |
Gas configuration | O2,N2O,Air |
Pipeline input range | 280-600Kpa |
Battery Power | |
Run-time | ≥180 mins |
Pre-setting Functions | |
Language | English, Chinese,etc. |
Calibration | Automatic |
Ventilator Specifications | |
Modes of Ventilation | |
VCV | Volume Controlled Ventilation |
A/C | Assist/Control Ventilation |
SIMV | Synchronized Intermittent Mandatory Ventilation |
PCV | Pressure Control Ventilation |
PSV | Pressure Support Ventilation |
Manual | Manual |
Ventilator Parameter | |
Working type | Electronic control, pneumatic driven |
Tidal Volume range | 20-1500ml |
Pressure range | 5-60cmH2O |
Rispriation rate range | 1-100bpm(1-40bpm under SIMV) |
I:E range | 4:1-1:8 |
Ptrigger range | -20-20 cmHz0 (based on PEEP) |
Flow trigger | 0-15L |
Minute volume range | >18L/Min |
SlGH range | 0-5/100 |
System standard | |
Hypoxic guard system | N2O cut-off valve, O2 concentration >25% |
Safety valve | <12.5kPa |
Fresh gas compensation | 25-75L/min |
Volume of CO2 absorber | 2 L |
Flow meter | 5 tubes, Cascade 0.1-15L/min O2 0.1-15L/min N2O, 0.1-15L/min air |
Vaporizer | 2 Selectetac, maximum 2 mounts(lsoflurane/Enflurane/Sevoflurane/Halothane) |
Monitoring | |
Display graphics | Waveforms of P-T, F-T, V-T, loops of P-V,V-F,F-P |
Types | VTi,VTe, MV, Pmax, Pplateau, Ri,Cs,Freg,FiO2, ETCO2 (Optional). |
Alarm Limits | |
Audio/Visual | No tidal volume, MV, Paw, FiO2 limitation O2,Air failure, AC power failure, Battery low |
PEEP | |
Type | Integrated, Electronic control |
Range | OFF. 3-30cmH2O |
Optional | |
1.AGSS | |
2. Suction Unit | |
3.Communication Interface |