Display
1.12.1 inch touch screen with no dead space.
2.New interface design with user friendly layout and colour assortment.
3.Day/Night display mode brings a better visual perception to users. Night mode suitable for dark environment.
4.Display a full range of ventilation modes including VCV, PCV, PSV, SIMV-V, SIMV-P, PA-VC(optional). Also display pressure, flow rate, volume and CO2 waveform in same interface.
5.Real-time detection for loops, airway pressure, flow rate, volume, compliance and airway resistance.
6.Comprehensive monitoring capabilities include ventilation parameters monitoring, preset parameters, alarm indication and breathing waveform monitoring.
7.Typical display for Zoom in/out, comparison, freeze.
8.Show last 48 hrs trends.
9.Up to 2000 logs.
Integrated Breathing Circuit
1.Integrated breathing circuit with APL valve, excellent tightness and easy to use.
2.Low essential resistance, good mechanical compliance.
3.Replacing drive bellow is not required when operating pediatric anesthesia.
4.Drain cup design ensure no water trap in the breathing circuit.
5.Built-up design duo flow sensor(no maintenance required).
6.2L Large volume of CO2 canister.
7.Autoclave 134 ℃。
High-precision Vaporizer
1.New vaporizer can be used for large flow anesthesia (15L/min).
2.Apply automatic compensation function for temperature, pressure and flow.
3.Apply Interlock Selectatec.
4.Agent type: Isoflurane, Enflurane, Halothane, Sevoflurane.
ACGO (Optional)
1.International standard design.(diameter:22mm)
2.Convenient for patients’ oxygen inhalation and postoperative recovery.
3.To connect with T circuit or bain circuit for open-operation.
AGSS (Optional)
1.Made by aluminum alloy. High strength, light weight, never rusts.
2.Enhance the safety of the environment in which members of staff in close proximity with waste anaesthetic gases and vapors ( agents) work.
Yoke system(Optional)
Up to 3 back up cylinders with A-type 11L size.
Technique Specifications | |
Physical Specifications | |
Dimensions | 70*85*140cm |
Weight | 65kg |
Casters | Front wheels with lock |
Screen | 12.1 inch LCD touchscreen |
Working Specifications | |
Power | AC 220V+/-10%,50Hz+/-2% |
Pipeline Supply | |
Gas configuration | O2,N2O,Air |
Pipeline input range | 280-600Kpa |
Battery Power | |
Run-time | ≥180 mins |
Pre-setting Functions | |
Language | English, Chinese, etc. |
Calibration | Automatic |
Applications | |
Patient range | Adult/Pediatric |
Ventilator Specifications | |
Modes of Ventilation | |
VCV | Volume Controlled Ventilation |
SIMV-V | Synchronized Intermittent Mandatory Ventilation-Volume |
SIMV-P | Synchronized Intermittent Mandatory Ventilation-Pressure |
MANUAL | Manua |
A/C | Assist/Control Ventilation |
PCV | Pressure Control Ventilation |
PSV | Pressure Control Ventilation |
STAND-BY | STAND-BY |
Ventilator Parameter | |
Working type | Electronic control, pneumatic driven |
Tidal Volume range | 20-1500ml |
Pressure range | 5-60cmH2O |
Rispriation rate range | 1-100bpm(1-40bpm under SIMV) |
I:E range | 4:1-1:8 |
Ptrigger range | -20-20cmH2O(based on PEEP) |
Flow trigger | 0-15L |
Minute volume range | >18L/Min |
SIGH range | 0-5/100 |
System Standard | |
Hypoxic guard system | N2O cut-off Valve, O2 concentration>25% |
Safety valve | <12.5Kpa |
Fresh gas compensation | 25-75L/mi |
Volume of CO2 absorber | 2L |
Flow meter | E-flowmeter; 0.1-15L/min O2; 0.1-15L/min NO2, 0.1-15L/min air |
Vaporizer | 2 Selectetac, maximum 2 mounts (Isoflurane/Enflurane/Sevoflurane/Halothane) |
Monitoring | |
Display graphics | Waveforms of P-T, F-T,V-T, loops of P-V, V-F, F-P(Zoom in/out, comparison, freeze) |
Types | VT, MV, BPM, Paw, Lung compliance, Inspiratory Platform, FiO2 |
Alarm Limits | |
Audio/Visual | No tidal volume, MV, Paw, FiO2 limitation O2,Air failure, AC power failure, Battery low. |
PEEP | |
Type | Integrated, Electronic control |
Range | OFF, 3-30cmH2O |
Optional | |
1.AGSS: Anesthetic Gas Scavenging Systems | |
2.Suction Unit | |
3.Communication Interface: Ethernet, USB, VGA | |
4.Auxiliary Oxygen meter |